4-4 Section 4-3 - General Procedure
4-3-2-2 Scanner Shutdown
1.) From the rear of the system. switch OFF the Circuit Breaker.
2.) Disconnect the Mains Power Cable by unfastening the metal wire fastener.
Figure 4-4 Circuit Breaker and Power Cable on Back of Scanner
NOTICE
Disconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the
uninterruptible power system (UPS) in the Back End Processor, forcing an ordered shutdown of the
system.
Circuit Breaker
Mains Power Cable
Cable Clip