GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-67
7-12-12 CD-RW Troubleshooting
Figure 7-57 CD-RW Troubleshooting
Unable to record to CD-R!
Insert a new, un-used CD-R disc.
YES
NO
Is a new, unused CD-R
disc inserted to the drive?
Replace BEP.
Are CD-Rs rated at the proper write
speed?
YES
NO
Acquire CD-Rs rated at the proper write
speed.