GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 8 Replacement Procedures 8-103
8-23-4 Console Lock Assembly Removal Procedure
1.) Unscrew 8 (5 mm) and 2 (3 mm) hexagon screws (See Figure 8-113).
2.) Take care of the spring.
3.) Remove the Console Lock Assembly.
8-23-5 Console Lock Assembly Installation Procedure
1.) Install the Lock Assembly.
2.) Install the spring.
3.) Fasten the 10 hexagon screws.
4.) Install Operator Panel.