GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 8 Replacement Procedures 8-159
Section 8-41
Frogleg Replacement Procedure
8-41-1 Manpower
2 persons, 3 hours
8-41-2 Tools
Hexagon key, 10 mm, with torque/momentum controller
Hexagon keys, 4, 5 and 6 mm
Phillips screwdriver, size 2 and 3
8-41-3 Preparations
1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
2.) Move Top Console to down position.
3.) Remove monitor as described in Monitor Assembly Replacement Procedure on page 8-92.
4.) Unscrew two (2) hexagon screws to loosen the Console Cable Neck from the Frogleg.
5.) Unscrew four (4) screws on the Upper Column Cover.