EasyManua.ls Logo

GE LOGIQ 9 - Unable to Scan

GE LOGIQ 9
570 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-57
7-12-4 Unable to Scan
Figure 7-50 Unable to Scan
Unable to scan!
Check the probe connection.
YES
NO
Are probes connected in
right way?
Restart the scanner.
Are other operations
possible?
YES
Check the HV power supply unit.
NO
END