8-19-5 Loudspeaker Installation Procedure
1.) Mount the new speaker assembly by loosely fastening the two top mounting screws and then the
two bottom mounting screws.
2.) Position the speaker assembly and tighten all four screws.
3.) Plug the speaker cable into the back of the monitor chassis.
4.) Replace the cover with speaker grills.
5.) Replace the two top screws on the cover but do not tighten them.
6.) Tilt the back of the monitor up and position it over the tray well to replace and tighten the two bottom
screws.
7.) Tighten the two top cover screws and replace the screw caps.
8.) Power On/Boot-up the scanner as described in Section 4-3-1 on page 4-2.
9.) Do functional check-out as described.
Figure 8-108 Speaker Mounting Screws