GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 8 Replacement Procedures 8-93
8-18-3 Preparations (click here to view the cover removal video) (cont’d)
e.) Slide the cover back and off of the monitor chassis. Ensure that both speaker grills are
removed.
NOTE: The two speaker grills fit loosely and may come off with the cover or remain with the monitor
chassis.
3.) Unplug the five (5) cables from the back of the monitor chassis. Remove the cables from any
holders or tie downs.
4.) Remove the two loudspeakers by removing the two bottom mounting screws from the speaker
assembly.
5.) Support the speaker assembly and remove the top mounting screws.
6.) Set the speakers aside.
Figure 8-103 Bottom Cover Screws