8-14 Section 8-3 - Back End Processor Replacement Procedure
8-3-4 Back-End Processor Removal Procedure
1.) Remove cables from the back of the Back-end Processor. As you remove each cable, ensure they
are properly marked. See Ta bl e 8 - 4 or Ta bl e 8 - 5 for cable identification.
2.) Loosen and remove the 6 fastening screws (4 at rear of processor, 2 at front) holding the processor
to the frame. Refer to Figure 8-9.
3.) Pull the Back-end Processor backwards and angle it to the outside to remove it from the frame.
Figure 8-9 Removing / Installing the Back-end Processor
Screws (6)