GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 6 Service Adjustments 6-5
6-3-2-5 Procedure
1.) Raise the Control Panel to a height that allows you to work underneath the monitor.
2.) Remove the eight (8) Phillips screws holding the Under Shelf Cover to the Shelf Cover Assembly.
3.) Remove four (4) Phillips screws holding the monitor Rear Cover Assembly to the monitor chassis.
4.) You now have access the monitor tilt and counter-balance adjustments. Adjust accordingly.
5.) Install the monitor rear cover assembly to the monitor chassis.
6.) Position the shelf cover assembly and secure it in place with the Under Shelf Cover and the eight
Phillips screws.
Section 6-4
Touch Panel Calibration
The Touch Panel Adjustment/Calibration procedure can be found in Section 7-7-4-10 “Calibration
Utilities” on page 7-34.