GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-17
When you select a test to run, like Troubleshoot all Acquisition Hardware, the display changes as shown
in Figure 7-18 on page 7-18 .
1.) Instructions Frame
- Displays either a test-specific text or the default instructions.
2.) Status Frame
- Initially displays the last known status for a selected diagnostic. Once the diagnostic starts,
the frame displays the “current” status of all test results.
- The Status Frame also contains the user interface elements used for Diagnostic Control and
Operator Feedback.
3.) Execute Button
- This switch has two modes - each with appropriate text:
- Execute - to start the diagnostic
- Abort - to stop a diagnostic