IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
8-188 Section 8-50 - Wireless Voice Scan Option
8-50-4-2Transmitter Adjustment (cont’d)
If you have multiple LOGIQ™ 9’s with wireless use the Arrow buttons to select a different preset and
push Set.
5.) Push Set again to enter the Menu and continue pushing Set until you get to Tune
6.) Select the frequency on the second LOGIQ™ 9’s receiver using the Arrow Buttons, push Set to
Enter.
7.) Check the RF LED on the receiver in the second LOGIQ 9, it should now be illuminated. The first
LOGIQ™ 9 RF LED should be off.
8.) Continue with the 3rd and 4th presets if additional LOGIQ™ 9’s are at the location.
9.) Perform the Volume Setup with each LOGIQ™ 9.