GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 10 Periodic Maintenance 10-3
Note 1: These specific tasks are completed during GE performed PMs or any time corrective
maintenance is performed by GE.
Note 2: May require specialized equipment to complete.
Table 10-2 Customer Performed Maintenance Tasks
Service at Indicated Time Daily Weekly Monthly
Per Facilities
QA Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders
•
Clean Air Filter •
more frequently depending on
your environment
Inspect AC Mains Cable • Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean Monitor and Touch Panel •
Inspect Wheels, Casters, brakes and
Swivel Locks
• Mobile Unit Check Daily
Check Control Panel Movement
• Mobile Unit Check Daily
Console Leakage Current Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Peripheral Leakage Current Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Surface Probe Leakage Current Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Endocavity Probe Leakage
Current Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Transesphongeal Probe Leakage
Current Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Surgical Probe Leakage
Current Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Measurement Accuracy Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Probe/Phantom Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program
Functional Checks
See Note 1
See Note 2
also after corrective
maintenance or as required by
your facilities QA program