GDM-906X Series User Manual
240
floating point with exponent
[MIN]
(Optional
parameter)
For commands, this will set the setting to the
lowest value. This parameter can be used in
place of any numerical parameter where
indicated.
For queries, it will return the lowest possible
value allowed for the particular setting.
[MAX]
(Optional
parameter)
For commands, this will set the setting to the
highest value. This parameter can be used in
place of any numerical parameter where
indicated.
For queries, it will return the highest possible
value allowed for the particular setting.
For commands, this will set the setting to the
default value. This parameter can be used in
place of any numerical parameter where
indicated.
For queries, it will return the default value
allowed for the particular setting.
Automatic
parameter
range
selection
The GDM-9060/9061 automatically sets the command
parameter to the next available value.
conf:volt:dc 3
This will set the measurement item to DC
Voltage and the range to 10V. There is no
3V range so the DMM selects the next
available range, 10V.
Marks the end of a command line. The
following messages are in accordance with
IEEE488.2 standard.
The most common
EOL character is
CR+LF