GDM-906X Series User Manual
358
AC Characteristics
[1]
True RMS AC Voltage
[2] [3] [4]
Range
[2]
Frequency
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
100 mV
3 Hz 5 Hz
1.00 + 0.03
1.00 + 0.04
1.00 + 0.04
0.100 + 0.004
5 Hz 10 Hz
0.38 + 0.03
0.38 + 0.04
0.38 + 0.04
0.035 + 0.003
10 Hz 20 kHz
0.07 + 0.03
0.08 + 0.04
0.09 + 0.04
0.005 + 0.003
20 kHz 50 kHz
0.13 + 0.04
0.14 + 0.05
0.15 + 0.05
0.011 + 0.005
50 kHz 100 kHz
0.58 + 0.08
0.63 + 0.08
0.63 + 0.08
0.060 + 0.008
100 kHz 300 kHz
4.00 + 0.50
4.00 + 0.50
4.00 + 0.50
0.200 + 0.020
1 V ~ 750 V
3 Hz 5 Hz
1.00 + 0.02
1.00 + 0.03
1.00 + 0.03
0.100 + 0.004
5 Hz - 10 Hz
0.38 + 0.02
0.38 + 0.03
0.38 + 0.03
0.035 + 0.003
10 Hz 20 kHz
0.07 + 0.02
0.08 + 0.03
0.09 + 0.03
0.005 + 0.003
20 kHz 50 kHz
0.13 + 0.04
0.14 + 0.05
0.15 + 0.05
0.011 + 0.005
50 kHz 100 kHz
0.58 + 0.08
0.63 + 0.08
0.63 + 0.08
0.060 + 0.008
100 kHz 300 kHz
4.00 + 0.50
4.00 + 0.50
4.00 + 0.50
0.200 + 0.020
Accuracy Specifications: ± ( % of reading + % of range )
True RMS AC Current
[2] [4] [5]
Range
[2]
Burden
Voltage
Frequency
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
100 μA/
10 mA
< 0.011 V,
< 0.04 V
3 Hz 5 Hz
1.00 + 0.04
1.00 + 0.04
1.00 + 0.04
0.100 + 0.006
5 Hz 10 Hz
0.38+ 0.04
0.38 + 0.04
0.38 + 0.04
0.035 + 0.006
10 Hz 5 kHz
0.13 + 0.04
0.13 + 0.04
0.13 + 0.04
0.015 + 0.006
5 kHz 10 kHz
0.20 + 0.04
0.20 + 0.04
0.20 + 0.04
0.030 + 0.006
1 mA/
100 mA
< 0.11 V,
< 0.4 V
3 Hz 5 Hz
1.00 + 0.04
1.00 + 0.04
1.00 + 0.04
0.100 + 0.006
5 Hz 10 Hz
0.33+ 0.04
0.33 + 0.04
0.33 + 0.04
0.035 + 0.006
10 Hz 5 kHz
0.13 + 0.04
0.13 + 0.04
0.13 + 0.04
0.015 + 0.006
5 kHz 10 kHz
0.18 + 0.04
0.18 + 0.04
0.18 + 0.04
0.030 + 0.006
1 A
< 0.7 V
3 Hz 5 Hz
1.00 + 0.04
1.00 + 0.04
1.00 + 0.04
0.100 + 0.006
5 Hz 10 Hz
0.33 + 0.04
0.33 + 0.04
0.33 + 0.04
0.035 + 0.006
10 Hz 5 kHz
0.13 + 0.04
0.13 + 0.04
0.13 + 0.04
0.015 + 0.006
5 kHz 10 kHz
0.18 + 0.04
0.18 + 0.04
0.18 + 0.04
0.030 + 0.006
3 A
< 2.0 V
3 Hz 5 Hz
1.00 + 0.04
1.00 + 0.04
1.00 + 0.04
0.100 + 0.006
5 Hz 10 Hz
0.38 + 0.04
0.38 + 0.04
0.38 + 0.04
0.035 + 0.006
10 Hz 5 kHz
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.015 + 0.006
5 kHz 10 kHz
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.030 + 0.006
Accuracy Specifications: ± ( % of reading + % of range )

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDM-906 Series and is the answer not in the manual?

GW Instek GDM-906 Series Specifications

General IconGeneral
BrandGW Instek
ModelGDM-906 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals