GDM-9061 Section
343
GDM-9061 Section
DC Characteristics
[1]
DC Voltage
Range
[2]
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
100.0000 mV
0.0030 + 0.0030
0.0040 + 0.0035
0.0050 + 0.0035
0.0005 + 0.0005
1.000000 V
0.0020 + 0.0006
0.0035 + 0.0007
0.0048 + 0.0007
0.0005 + 0.0001
10.00000 V
0.0015 + 0.0004
0.0020 + 0.0005
0.0035 + 0.0005
0.0005 + 0.0001
100.0000 V
0.0020 + 0.0006
0.0035 + 0.0006
0.0050 + 0.0006
0.0005 + 0.0001
1000.000 V
0.0025 + 0.0006
0.0040 + 0.0010
0.0050 + 0.0010
0.0005 + 0.0001
Accuracy Specifications: ± ( % of reading + % of range )
Resistance
[3]
Range
[2]
Test
Current
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
100.0000
1 mA
0.003 + 0.0030
0.008 + 0.004
0.010 + 0.004
0.0008 + 0.0005
1.000000 k
1 mA
0.002 + 0.0005
0.008 + 0.001
0.010 + 0.001
0.0008 + 0.0001
10.00000 k
100 μA
0.002 + 0.0005
0.008 + 0.001
0.010 + 0.001
0.0008 + 0.0001
100.0000 k
10 μA
0.002 + 0.0005
0.008 + 0.001
0.010 + 0.001
0.0008 + 0.0001
1.000000 M
5 μA
0.002 + 0.0010
0.008 + 0.001
0.010 + 0.001
0.0010 + 0.0002
10.00000 M
500 nA
0.015 + 0.0010
0.020 + 0.001
0.040 + 0.001
0.0030 + 0.0004
100.0000 M
500 nA//
10 M
0.300 + 0.0100
0.800 + 0.010
0.800 + 0.010
0.1500 + 0.0002
Accuracy Specifications: ± ( % of reading + % of range )
DC Current
Range
[2]
Burden
Voltage
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
100.0000 μA
< 0.011 V
0.010 + 0.020
0.040 + 0.025
0.050 + 0.025
0.002 + 0.003
1.000000 mA
< 0.11 V
0.007 + 0.006
0.030 + 0.006
0.050 + 0.006
0.002 + 0.001
10.00000 mA
< 0.04 V
0.007 + 0.020
0.030 + 0.020
0.050 + 0.020
0.002 + 0.002
100.0000 mA
< 0.4 V
0.010 + 0.004
0.030 + 0.005
0.050 + 0.005
0.002 + 0.001
1.000000 A
< 0.7 V
0.050 + 0.006
0.080 + 0.010
0.100 + 0.010
0.005 + 0.001
3.000000 A
< 2.0 V
0.180 + 0.020
0.200 + 0.020
0.200 + 0.020
0.005 + 0.002
10.00000 A
[6]
< 0.5 V
0.100 + 0.010
0.120 + 0.010
0.150 + 0.010
0.005 + 0.001
Accuracy Specifications: ± ( % of reading + % of range )
Continuity
Range
[2]
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
1 kΩ
0.002 + 0.030
0.008 + 0.030
0.01 + 0.03
0.001 + 0.002
Accuracy Specifications: ± ( % of reading + % of range )

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDM-906 Series and is the answer not in the manual?

GW Instek GDM-906 Series Specifications

General IconGeneral
BrandGW Instek
ModelGDM-906 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals