GDM-906X Series User Manual
256
Other Commands
ABORt
Aborts a measurement in progress, returning the instrument to the trigger
idle state.
Use this to abort a measurement when the instrument is waiting for a
trigger, or for aborting a long measurement or series of measurements.
FETCh[X]?
Waits for measurements to complete and copies all available
measurements to the instrument's output buffer. The readings remain in
reading memory.
X = null or 1 indicate 1
st
display value, X = 2 indicate 2
nd
display value
Example: SAMP:COUN 3
INIT
FETC?
Returns: -4.98748741E-01,-4.35163427E-01,-4.33118686E-01
The FETCh? query does not erase measurements from the reading
memory. You can send the query multiple times to retrieve the same data.
You can store up to 10,000 measurements in the reading memory of the
GDM-9060 or 100,000
measurements on the GDM-9061. If reading memory overflows,new
measurements overwrite the oldest measurements stored; the most recent
measurements are always preserved. No error is generated, but the Reading
Mem Ovfl bit (bit 14) is set in the Questionable Data Register's condition
register.
HCOPy:SDUMp:DATA?
Executes TFT LCD screenshot action.
Returns the front panel display image ("screen shot").
Returns a count of data streaming by the image file format of BMP.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDM-906 Series and is the answer not in the manual?

GW Instek GDM-906 Series Specifications

General IconGeneral
BrandGW Instek
ModelGDM-906 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals