MEASure:CAPacitance
Returns the capcitance measurement value on the 1
st
display
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:CAP?
Returns the capcitance measurement value.
MEASure:CONTinuity?
Returns the continuity measurement value on the 1
st
display.
Example: MEAS:CONT?
Returns the continuity measurement value.
MEASure:DIODe?
Returns the diode measurement value on the 1
st
display.
Example: MEAS:DIOD?
Returns the diode measurement value.
MEASure:TEMPerature?
Returns the temperature measurement value with the selected probe and
type on the 1
st
display.
Parameter: [None] | [Probe type, Type, 1, Resolution(<NRf> | MIN | MAX |
DEF)]
< Probe type >:
TCOuple | RTD | FRTD | THERmistor | FTHermistor
<Type>:
Tcouple: J | K | N | R | S | T | B | E
RTD / FRTD : PT100 | D100 | F100 | PT385 | PT3916 | USER
Thermistor / Fthermistor : 2.2k | 5k | 10k | USER
Example: MEAS:TEMP? TCO,K
> +0.26561348E+02
Returns the temperature measurement value.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
Secondary Display: MEASure2 Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDM-906 Series and is the answer not in the manual?

GW Instek GDM-906 Series Specifications

General IconGeneral
BrandGW Instek
ModelGDM-906 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals