REMOTE CONTROL
275
MEASure:CURRent:AC?
Returns the AC current measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:CURR:AC?
> +1.46445157E-07
Returns the AC current measurement value as 0.000146 mA.
MEASure:RESistance?
Returns the 2-wire resistance measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF),
Resolution(<NRf> | MIN | MAX | DEF)]
Example: MEAS:RES? 1,MIN
> +1.18137284E+06
Sets measurement mode to 2-wire resistance with a range of 1 at the
mimimum resolution and return measurement value.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
MEASure:FRESistance?
Returns the 4-wire resistance measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF),
Resolution(<NRf> | MIN | MAX | DEF)]
Example: MEAS:FRES?
> +1.18134472E+06
Sets measurement mode to 4-wire resistance to auto range and return
measurement value.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
MEASure:FREQuency?
Returns the frequency measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:FREQ?
> +0.21504529E+05
Returns the frequency measurement value as 21.5 kHz.
MEASure:PERiod?
Returns the period measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:PER? MAX
Returns the period measurement value at the maximum range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDM-906 Series and is the answer not in the manual?

GW Instek GDM-906 Series Specifications

General IconGeneral
BrandGW Instek
ModelGDM-906 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals