EasyManua.ls Logo

GW Instek GDM-906 Series - Page 275

GW Instek GDM-906 Series
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
275
MEASure:CURRent:AC?
Returns the AC current measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:CURR:AC?
> +1.46445157E-07
Returns the AC current measurement value as 0.000146 mA.
MEASure:RESistance?
Returns the 2-wire resistance measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF),
Resolution(<NRf> | MIN | MAX | DEF)]
Example: MEAS:RES? 1,MIN
> +1.18137284E+06
Sets measurement mode to 2-wire resistance with a range of 1 at the
mimimum resolution and return measurement value.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
MEASure:FRESistance?
Returns the 4-wire resistance measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF),
Resolution(<NRf> | MIN | MAX | DEF)]
Example: MEAS:FRES?
> +1.18134472E+06
Sets measurement mode to 4-wire resistance to auto range and return
measurement value.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
MEASure:FREQuency?
Returns the frequency measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:FREQ?
> +0.21504529E+05
Returns the frequency measurement value as 21.5 kHz.
MEASure:PERiod?
Returns the period measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:PER? MAX
Returns the period measurement value at the maximum range.

Table of Contents