[SENSe:]CONTinuity:TRIGger:DELay
Sets the trigger delay that minimum step is microseconds of Continuity
measurement.
Parameter: <NRf>(0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:CONT:TRIG:DEL 0.0001
Sets the trigger delay time to 100us of Continuity measurement.
[SENSe:]CONTinuity:TRIGger:DELay?
Returns the trigger delay time in seconds of Continuity measurement.
Return parameter: <NRf>
[SENSe:]CONTinuity:ZERO:AUTO
Sets the auto zero mode to on, off or once only of Continuity measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:CONT:ZERO:AUTO OFF
Sets the auto zero to off.
[SENSe:]CONTinuity:ZERO:AUTO?
Returns the auto zero mode of Continuity measurement.
Return parameter: 0 | 1, 1=ON, 0=OFF
SENSe DIODe Commands
[SENSe:]DIODe:NPLCycles
Sets the integration time for Diode measurement in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.15 | 0.6 | 1).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:DIOD:NPLC DEF
Sets the integration time to the 1 PLCs for diode measurement.
[SENSe:]DIODe:NPLCycles?
Returns the integration time for Diode measurement in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.15 | 0.6 | 1
[SENSe:]DIODe:RESolution
Sets the Diode measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example:SENS:DIOD:RES 0.1e-4
Sets the Diode resolution to 0.00001
[SENSe:]DIODe:RESolution?
Returns the Diode measurement resolution.