REMOTE CONTROL
289
[SENSe:]CURRent[:DC]:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of DC
Current measurement.
Parameter: <NRf>(0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:CURR:DC:TRIG:DEL 2e-4
Sets the trigger delay time to 200us of DC Current measurement.
[SENSe:]CURRent[:DC]:TRIGger:DELay?
Returns the trigger delay time in seconds of DC Current measurement.
Return parameter: <NRf>
[SENSe:]CURRent[:DC]:ZERO:AUTO
Sets the auto zero to on, off or once of DC Current measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:CURR:DC:ZERO:AUTO ON
Sets the auto zero to on.
[SENSe:]CURRent[:DC]:ZERO:AUTO?
Returns the auto zero mode of DC Current measurement.
Return Parameter: 0 | 1, 1=ON, 0=OFF
[SENSe:]CURRent:AC:BANDwidth
Sets the AC current bandwidth (AC filter).
Parameter: <NRf> (3 | 20 | 200) | MIN | MAX | DEF
Example: SENS:CURR:AC:BAND 3
Sets the AC current bandwidth to 3Hz.
[SENSe:]CURRent:AC:BANDwidth?
Returns the AC current bandwidth.
Return parameter: <NRf>
[SENSe:]CURRent:AC:NULL[:STATe]
Sets the relative function on/off for AC Current measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:CURR:AC:NULL:STAT ON
Turns the relative function on for AC Current measurement.
[SENSe:]CURRent:AC:NULL[:STATe]?
Returns the relative function state of AC Current measurement.
Return parameter: 0|1, 0=OFF, 1=ON
[SENSe:]CURRent:AC:NULL:VALue
Sets the relative value for AC Current measurement.
Parameter: <NRf> (-12.0~12.0 A) | MIN | MAX | DEF
Example: SENS:CURR:AC:NULL:VAL 0.02
Sets the relative value to 0.02A for AC Current measurement.