REMOTE CONTROL
307
TRIGger Commands
SAMPle:COUNt
Sets the number of samples.
Parameter: <NRf>(1.0 ~ 1000000.0) | MIN | MAX | DEF
Example: TRIG:COUN 10
SAMP:COUN 10
INIT
FETC?
Will returns 100 measurment results.
Sets the number of samples to 10.
The total measurement counts is trigger count multiplication sample
count.
SAMPle:COUNt?
Returns the number of samples.
Return parameter: <NRf>
TRIGger:COUNt
Sets the number of trigger counts.
Parameter: <NRf>(1.0 ~ 1000000.0) | MIN | MAX | DEF
Example: TRIG:COUN 10
SAMP:COUN 10
READ?
Will returns 100 measurment results.
Sets the number of trigger counts to 10.
The total measurement counts is trigger count multiplication sample
count.
TRIGger:COUNt?
Returns the number of trigger counts.
Return parameter: <NRf>
TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds in all of the
function.
Parameter: <NRf> (0 ~ 3600 s) | MIN | MAX | DEF
Example: TRIG:DEL MAX
Sets the trigger delay time to the maximum.
TRIGger:DELay?
Returns the trigger delay time in seconds of current function.
Return parameter: <NRf>