REMOTE CONTROL
329
The following table lists the bit definitions for the Operation Data Register:
Bit
Name
Decimal
Definition
0
Calibrating
1
Instrument is performing a calibration.
1
Not Used
2
(Reserved for future use)
2
Not Used
4
(Reserved for future use)
3
Not Used
8
(Reserved for future use)
4
Measuring
16
Instrument is initiated, and is making or about to make
a measurement.
5
Waitig For
Trigger
32
Instrument is waiting for a trigger.
6
Not Used
64
(Reserved for future use)
7
Not Used
128
(Reserved for future use)
8
Configurtion
Change
256
Instrument configuration has been changed since the
last INIT, READ? or MEASure?, either from the front
panel or from SCPI.
9
Data Output
Threshold
512
Programmed number of measurements
(DATA:POINts:EVENt:THReshold) have been stored
in measurement memory.
10
Not Used
1024
(Reserved for future use)
11
Not Used
2048
(Reserved for future use)
12
Not Used
4096
(Reserved for future use)
13
Global Error
8192
Set if any remote interface has an error in its error
queue; cleared otherwise.
14
Not Used
16384
(Reserved for future use)
15
Not Used
32768
(Reserved for future use)