BASIC MEASUREMENT
63
3. Press the F6 (Enter) key or the Knob key to
confirm the input α value and repeat the previous steps 2 - 4
to set up the B and C coefficients individually.
B range: 0 ~ 9.9999 (default :2.35600E-04)
C range: 0 ~ 9.9999 (default :9.55700E-08)
THERM B Setup
THERM C Setup
4. After returning to the User Type Setup page, if necessary,
press the F6 (5 kΩ DEF) key to restore to the
default coefficients’ setting based on the 5 kΩ sensor type.