SAFETY INSTRUCTIONS
7
WARNING
(GDM-9061 only)
Due to the fact that the Front/Rear Input Switch on
the front panel is not proposed as an active
multiplexer, do Not change the input switch when
signals are present on either rear or front set of
terminals. Intrument damage and risk of electric shock
may occur if switching the input switch when high
voltage or current is present.
Fuse
WARNING
Fuse type: T0.25A 100/120 VAC
T0.125A 220/240 VAC
Make sure the correct type of fuse is installed before
power up.
To avoid risk of fire, replace the fuse only with the
specified type and rating.
Disconnect the power cord before fuse replacement.
Make sure the cause of a fuse blowout is fixed before
fuse replacement.
Cleaning the
Instrument
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into the
GDM-9060/9061.
Do not use chemicals or cleaners containing harsh
material such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free, almost
non-conductive pollution (Note below)
Temperature: Full accuracy for 0°C to 55°C.
Humidity:
< 30°C: < 80%RH (non-condensing)
30°C~40°C: <70%RH (non-condensing)
>40°C: <50%RH (non-condensing)
Altitude: <2000m

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDM-906 Series and is the answer not in the manual?

GW Instek GDM-906 Series Specifications

General IconGeneral
BrandGW Instek
ModelGDM-906 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals