COMMAND DETAILS
105
:MEASure:PHAse
Description
Returns the phase between source 1 and source 2.
Syntax
:MEASure:PHAse{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the phase difference.
Chan Off
Indicates the source channel is not
activated.
Note
Select the two source channels before entering this
command.
Example
:MEASure:SOURce1 CH1
:MEASure:SOURce2 CH2
:MEASure:PHAse?
4.50E+01
Select channel 1 and 2 as phase source1/2, and
then measure the phase in degrees.