GDS-1000B Programming Manual
152
:TRIGger:BUS:B1:SPI:DATa:MOSI:VALue?
1X1X0101
:TRIGger:BUS:B1:CAN:CONDition
Sets or returns the CAN trigger condition.
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
:TRIGger:BUS:B1:CAN:CONDition
{SOF|FRAMEtype|IDentifier|DATA|IDANDDATA|EOF|
ACKMISS|STUFFERR|?}
Parameter/
Return parameter
Triggers on a start of frame
Triggers on the type of frame
Triggers on a matching identifier
Triggers on matching data
Triggers on matching identifier and
data field
Triggers on the end of frame
Triggers on a missing acknowledge
Triggers on a bit stuffing error
:TRIGger:BUS:B1:CAN:CONDition SOF
Triggers on a start of frame.
:TRIGger:BUS:B1:CAN:CONDition?
>SOF
:TRIGger:BUS:B1:CAN:FRAMEtype
Sets or returns the frame type for a CAN
FRAMEType trigger.