Data Logging Commands
The data logging commands are only applicable after the optional
Data Log app has been installed. Please see the user manual for
further details.
:DATALOG:STATE ..................................................... 244
:DATALOG:SOURce .................................................. 244
:DATALOG:SAVe ....................................................... 245
:DATALOG:INTerval .................................................. 245
:DATALOG:DURation ................................................ 246
Sets or queries the state of the data logging app.
:DATALOG:STATE {OFF|ON|?}
:DATALOG:SOURce
:DATALOG:SAVe
:DATALOG:INTerval
:DATALOG:DURation
Parameter/
Return parameter
Turns the data logging off.
Turns the data logging on.
:DATALOG:STATE ON
Turns the data logging app on.
Sets or queries the data logging source channel.
:DATALOG:SOURce {CH1~CH4|all|?}