EasyManua.ls Logo

GW Instek GDS-1000B Series - Math:fft:mag; Math:fft:source; Math:fft:window

GW Instek GDS-1000B Series
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND DETAILS
57
:MATH:FFT:SOURce
Description
Sets and queries the FFT math source.
Syntax
:MATH:FFT:SOURce { CH1 | CH2 | CH3 | CH4 | REF1 |
REF2 | REF3 | REF4 | ? }
Related
commands
:MATH:ADVanced:EDIT:SOURce<X>
:MATH:ADVanced:EDIT:OPERator
Parameter
CH1~4
Channel 1 to 4
REF1~4
Reference waveform 1 to 4
Return parameter
Returns the FFT source.
Example
:MATH:FFT:SOURce CH1
Sets the FFT math source as channel 1.
:MATH:FFT:MAG
Description
Sets FFT vertical units as linear or decibels.
Syntax
:MATH:FFT:MAG {LINEAR | DB | ?}
Parameter
LINEAR
Linear units (Vrms)
DB
Logarithmic units (dB)
Return parameter
Returns the FFT vertical units.
Example
:MATH:FFT:MAG DB
Sets FFT vertical units to dB.
:MATH:FFT:WINDow
Description
Sets the windowing filter used for the FFT
function.
Syntax
:MATH:FFT:WINDow
{RECTangular|HAMming|HANning|BLAckman|?}
Set
Query
Set
Query
Set
Query