GDS-1000B Programming Manual
232
Example
:SEGMents:STATE ON
STOP
:SEGMents:MEASure:MODe PLOT
:SEGMents:MEASure:PLOT:SOURce 1
:SEGMents:MEASure:PLOT:DIVide 10
:SEGMents:MEASure:PLOT:SELect 1
:SEGMents:MEASure:PLOT:RESults?
> MAX,1.000kHz;MIN,1.000kHz;MEAN,1.000kHz;
Bin Statistics,1 of 10;Percent,10.00%;Count,1;
Measured,10;Unmeasured,0;Bin Range,
1.000kHz~1.000kHz;
Plots the results for automatic measurement #1,
bin 1 of 10.
:SEGMents:MEASure:TABle:SOURce
Description
Sets or queries the source of the measurement list.
Syntax
:SEGMents:MEASure:TABle:SOURce {CH1 | CH2 |
CH3| CH4 | ? }
Related
commands
:SEGMents:MEASure:MODe
:SEGMents:MEASure:TABle:SELect
:SEGMents:MEASure:TABle:LIST
Parameter/
Return parameter
CH1~CH4
Channel 1 to 4
Example1
:SEGMents:MEASure:TABle:SOURce CH1
Sets the source to CH1.
:SEGMents:MEASure:TABle:SELect
Description
Sets or queries a segment to view in the
measurement table.
Syntax
:SEGMents:MEASure:TABle:SELect {<NR1> | ? }
Set
Query
Set
Query