COMMAND DETAILS
39
7
128
Not used
Example
*SRE?
>48
Indicates that the MAVB and ESB bit are both set.
*STB
Description
Queries the bit sum of the Status Byte register with
MSS (Master summary Status) replacing the RQS
bit (bit 6).
Query Syntax
*STB?
Return parameter
<NR1>
0 ~ 255
Bit Weight
Bit#
Weight
Event
Description
0
1
Not used
1
2
Not used
2
4
Not used
3
8
Not used
4
16
MAV
Message Available Bit
5
32
ESB
Event Status Bit
6
64
MSS
Master Summary Bit
6
64
RQS
Request Service Bit
7
128
Not used
Example
*STB?
>16
Indicates that the MAV bit is set.