EasyManua.ls Logo

GW Instek GDS-1000B Series - Cursor:hdelta; Cursor:v1 Position; Cursor:v2 Position

GW Instek GDS-1000B Series
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDS-1000B Programming Manual
68
:CURSor:HDELta
Description
Returns the delta of H1 and H2.
Syntax
:CURSor:HDELta{?}
Return Parameter
<NR3>
Returns the distance between two
horizontal cursors.
Example
:CURSor:HDELta?
5.0E-9
Returns the horizontal delta as 5ns.
:CURSor:V1Position
Description
Sets the first vertical cursor (V1) position.
Syntax
:CURSor:V1Position {<NRf>| ?}
Parameter
<NRf>
Vertical position. Depends on the
vertical scale.
Return parameter
<NR3>
Returns the cursor position.
Example
:CURSor:V1Position 1.6E -1
Sets the V1 cursor position to 160mA.
:CURSor:V2Position
Description
Sets the first vertical cursor (V2) position.
Syntax
:CURSor:V2Position {<NRf> | ?}
Parameter
<NRf>
Vertical position. Depends on the
vertical scale.
Return parameter
<NR3>
Returns the cursor position.
Example
:CURSor:V2Position 1.1E-1
Sets the V2 cursor position to 110mA.
Query
Set
Query
Set
Query

Table of Contents