GDS-1000B Programming Manual
150
Note
This setting only applies when the SPI trigger is set
to trigger on MISO, MOSI or MISO/MOSI.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:SPI:DATa:SIZe {<NR1> | ?}
Related
commands
:TRIGger:BUS:B1:SPI:CONDition
Parameter
<NR1>
Number of words (1 to 32).
Return parameter
<NR1>
Returns the number of words.
Example
:TRIGger:BUS:B1:SPI:DATa:SIZe 10
Sets the number of words to 10.
:TRIGger:BUS:B1:SPI:DATa:MISO:VALue
Description
Sets or queries the triggering data value for the SPI
bus when the bus is set to trigger on MISO or
MISO/MOSI.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:SPI:DATa:MISO:VALue {<string>
| ? }
Related
commands
:TRIGger:BUS:B1:SPI:DATa:SIZe
Parameter
<sting>
The number of characters in the string
depends on the data size setting. The
string must be enclosed in double
quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the data value.