GDS-1000B Programming Manual
146
Parameter
RXSTArt
Set trigger on the RX Start Bit.
RXDATA
Set trigger on RX Data.
RXENDPacket
Set trigger on the RX End of Packet
condition.
RXPARItyerr
Set trigger on RX Parity error
condition.
TXSTArt
Set trigger on the TX Start Bit.
TXDATA
Set trigger on TX Data.
TXENDPacket
Set trigger on the TX End of Packet
condition.
TXPARItyerr
Set trigger on TX Parity error
condition.
Return Parameter
Returns the triggering condition.
Example
:TRIGger:BUS:B1:UART:CONDition TXDATA
Sets the UART bus to trigger on Tx Data.
:TRIGger:BUS:B1:UART:RX:DATa:SIZe
Description
Sets or queries the number of bytes for UART data.
Note
This setting only applies when the UART trigger is
set to trigger on Rx Data
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:UART:RX:DATa:SIZe {<NR1> | ?}
Related
commands
:TRIGger:BUS:B1:UART:CONDition
Parameter
<NR1>
Number of bytes (1 to 10).
Return parameter
<NR1>
Returns the number of bytes.
Example
:TRIGger:BUS:B1:UART:RX:DATa:SIZe 5
Sets the number of bytes to 5.
Set
Query