COMMAND DETAILS
153
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:CAN:FRAMEtype
{DATA|REMote|ERRor|OVERLoad|?}
Parameter/
Return parameter
DATA
Sets the frame type to data frame
REMote
Sets the frame type to remote frame
ERRor
Sets the frame type to error frame
OVERLoad
Sets the frame type to overload
Example
:TRIGger:BUS:B1:CAN:FRAMEtype DATA
Sets the frame type to DATA.
:TRIGger:BUS:B1:CAN:IDentifier:MODe
Description
Sets or returns the CAN identifier mode for the
bus.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:CAN:IDentifier:MODe
{STANDard|EXTended|?}
Parameter/
Return parameter
STANDard
Standard addressing mode
EXTended
Extended addressing mode
Example
:TRIGger:BUS:B1:CAN:IDentifier:MODe?
>STANDARD
Returns the addressing mode.
:TRIGger:BUS:B1:CAN:IDentifier:VALue
Description
Sets or returns the identifier string used for the
CAN trigger.
Note: Only applicable when the trigger condition
is set to ID or IDANDDATA.
Set
Query
Set
Query