COMMAND DETAILS
151
Example1
:TRIGger:BUS:B1:SPI:CONDition MISO
:TRIGger:BUS:B1:SPI:DATa:SIZe 2
:TRIGger:BUS:B1:SPI:DATa:MISO:VALue "1x1x0101"
Sets the value to 1x1x0101
Example 2
:TRIGger:BUS:B1:SPI:DATa:MISO:VALue?
1X1X0101
:TRIGger:BUS:B1:SPI:DATa:MOSI:VALue
Description
Sets or queries the triggering data value for the SPI
bus when the bus is set to trigger on MOSI or
MISO/MOSI.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:SPI:DATa:MOSI:VALue {<string>
| ? }
Related
commands
:TRIGger:BUS:B1:SPI:DATa:SIZe
Parameter
<sting>
The number of characters in the string
depends on the data size setting. The
string must be enclosed in double
quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the data value.
Example1
:TRIGger:BUS:B1:SPI:CONDition MOSI
:TRIGger:BUS:B1:SPI:DATa:SIZe 2
:TRIGger:BUS:B1:SPI:DATa:MOSI:VALue "1x1x0101"
Sets the value to 1x1x0101
Set
Query