GDS-1000B Programming Manual
176
Parameter/Return
parameter
<NR1>
0: NULL
1: LF (line feed)
2: CR (carriage return)
3: SP (space character)
4: FF
Example
:BUS1:UART:EOFPAcket 2
Sets the OEF character to CR.
:BUS1:UART:TX:SOURce
Description
Sets or queries which channel is used for the
UART Tx source.
Syntax
:BUS1:UART:TX:SOURce { OFF | CH1 | CH2 | CH3 |
CH4 | ? }
Parameter/Return
parameter
OFF
Off, no Tx source
CH1 to CH4
Analog channels CH1 to CH4
Example
:BUS1:UART:TX:SOURce CH1
Sets channel 1 as the Tx source.
:BUS1:UART:RX:SOURce
Description
Sets or queries which channel is used for the
UART Rx source.
Syntax
:BUS1:UART:RX:SOURce { OFF | CH1 | CH2 | CH3 |
CH4 | ? }
Parameter/Return
parameter
OFF
Off, no Rx source
CH1 to CH4
Analog channels CH1 to CH4
Example
:BUS1:UART:RX:SOURce CH1
Sets channel 1 as the Rx source.