COMMAND DETAILS
37
4
16
EXE
Execution Error
5
32
CME
Command Error
6
64
URQ
User Request
7
128
PON
Power On
Example
*ESE?
>4
Indicates that there is a query error.
*ESR
Description
Queries the Standard Event Status (Event) register.
The Event Status register is cleared after it is read.
Query Syntax
*ESR?
Return parameter
<NR1>
0~255
Bit Weight
Bit#
Weight
Event
Description
0
1
OPC
Operation Complete Bit
1
2
RQC
Not used
2
4
QYE
Query Error
3
8
DDE
Device Error
4
16
EXE
Execution Error
5
32
CME
Command Error
6
64
URQ
User Request
7
128
PON
Power On