GDS-1000B Programming Manual
88
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:PDUTy?
5.000E+01
Selects Channel 1, and then measures the positive
duty cycle.
:MEASure:PERiod
Description
Returns the period.
Syntax
:MEASure:PERiod{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the period.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:PERiod?
1.0E-3
Selects Channel 1, and then measures the period.
:MEASure:PWIDth
Description
Returns the first positive pulse width.
Syntax
:MEASure:PWIDth{?}
Related
Commands
:MEASure:SOURce<X>
Query
Query