:MEASUrement:MEAS<X>:MEAN
Description
Returns the mean measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MEAN?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
Return parameter
<NR3>
Returns the measurement for the
selected measurement number.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MEAN?
2.090E-02
Returns the mean measurement result for
measurement number 3.
:MEASUrement:MEAS<X>:MINImum
Description
Returns the minimum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MINImum?
Related
Commands
:MEASUrement:STATIstics:MODe
Query