EasyManua.ls Logo

GW Instek GDS-1000B Series - Measurement:statistics:mode; Measurement:statistics:weighting

GW Instek GDS-1000B Series
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDS-1000B Programming Manual
112
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:STDdev?
1.530E-03
Returns the standard deviation for measurement
number 3.
:MEASUrement:STATIstics:MODe
Description
Puts the statics measurement results on the display
or queries whether the statistics are displayed.
Syntax
:MEASUrement:STATIstics:MODe {OFF | ON | ?}
Related
commands
:MEASUrement:STATIstics
Parameter/
Return parameter
ON
Display the statistics on the screen.
OFF
Remove the statistics from the screen
Example
:MEASUrement:STATIstics:MODe ON
Displays statistics on the screen.
:MEASUrement:STATIstics:WEIghting
Description
Sets and queries the number of samples
(weighting) used for the statistics calculations.
Syntax
:MEASUrement:STATIstics:WEIghting { <NR1> | ? }
Parameter/
Return parameter
<NR1>
Number of samples (2~1000)
Example
:MEASUrement:STATIstics:WEIghting 5
Sets the number of samples to 5.
Set
Query
Set
Query

Table of Contents

Other manuals for GW Instek GDS-1000B Series

Related product manuals