COMMAND DETAILS
121
Example
:TIMebase:WINDow:POSition 2.0E-3
Sets the zoom horizontal position as 20ms.
:TIMebase:WINDow:SCALe
Description
Sets or queries the zoom horizontal scale.
Note
If the oscilloscope is under “ZOOM” mode, the
main timebase function will be disabled and
cannot be modified.
Syntax
:TIMebase:WINDow:SCALe {<NRf> | ?}
Related
commands
:TIMebase:MODe
Parameter
<NRf>
Zoom horizontal scale. The range will
depend on the time base.
Return parameter
<NR3>
Returns the zoom horizontal scale.
Example
:TIMebase:WINDow:SCALe 2.0E-3
Sets the zoom horizontal scale to 2ms.
Set
Query