GDS-1000B Programming Manual
158
Parameter/
Return parameter
SYNCField
Sets the LIN trigger condition to the
sync field.
IDentifier
Sets the LIN trigger condition to
identifier field.
DATA
Sets the LIN trigger condition to the
data field.
IDANDDATA
Sets the LIN trigger condition to
identifier and data field
WAKEup
Sets the LIN trigger condition to wake
up.
SLEEP
Sets the LIN trigger condition to
sleep.
ERRor
Sets the LIN trigger condition to
error.
Example
:TRIGger:BUS:B1:LIN:CONDition?
>IDANDDATA
:TRIGger:BUS:B1:LIN:CONDition DATA
:TRIGger:BUS:B1:LIN:CONDition?
>DATA
:TRIGger:BUS:B1:LIN:DATa:QUALifier
Description
Sets or returns the LIN data qualifier.
Note: Only applicable when the trigger condition
is set to DATA or IDANDDATA.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:LIN:DATa:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSEQual|M
OREEQual|?}
Set
Query