COMMAND DETAILS
191
Example
:SEARCH:TRIGger:EDGe:SLOP FALL
Sets the search trigger slope to falling.
:SEARCH:TRIGger:LEVel
Description
Sets or queries the search trigger level.
Syntax
:SEARCH:TRIGger:LEVel {TTL | ECL| SETTO50 |
<NRf> | ?}
Related
commands
:SEARCH:TRIGger:TYPe
Parameter
<NRf>
Trigger level value
TTL
Sets the search trigger level to TTL.
ECL
Sets the search trigger level to ECL.
SETTO50
Sets the search trigger level to the
User level (50% by default).
Return parameter
<NR3>
Returns the trigger.
Example1
:SEARCH:TRIGger:LEVel TTL
Sets the search trigger level to TTL.
Example2
:SEARCH:TRIGger:LEVel 3.30E-1
Sets the search trigger level to 330mV/mA.
:SEARCH:TRIGger:HLEVel
Description
Sets or queries the high level search trigger.
Note
Applicable for Rise and Fall/Pulse Runt search
triggers.
Syntax
:SEARCH:TRIGger:HLEVel { <NRf> | ?}
Related
commands
:SEARCH:TRIGger:TYPe
Parameter
<NRf>
High level value.
Return parameter
<NR3>
Returns the high level search trigger.