COMMAND DETAILS
109
Example
:MEASUrement:MEAS1:SOUrce1 CH1
:MEASUrement:MEAS1:TYPe PK2PK
:MEASUrement:MEAS1:STATE ON
:MEASUrement:MEAS1:VALue?
5.000E+0
Selects channel 1 as the source for measurement 1,
sets measurement 1 to peak to peak measurement
and then turns on the measurement. The result
returns the peak to peak measurement.
:MEASUrement:MEAS<X>:MAXimum
Description
Returns the maximum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MAXimum?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
Return parameter
<NR3>
Returns the measurement for the
selected measurement number.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MAXimum?
2.800E-02
Returns the maximum measurement result for
measurement number 3.
Query