COMMAND DETAILS
55
:MATH:TYPe
Description
Queries or sets the Math type to FFT, Advanced
Math or to dual channel math operations (basic +, -
, ×, ÷, operations).
Syntax
:MATH:TYPe { DUAL | ADVanced | FFT | ? }
Related
Commands
:MATH:DISP
Parameter
DUAL
Dual channel operations
ADVanced
Advanced math operations
FFT
FFT operations
Return parameter
Returns the math type.
Example
:MATH:TYPe DUAL
Sets the Math type to dual channel math operation.
:MATH:DUAL:SOURce<X>
Description
Sets the dual math source for source 1 or 2.
Syntax
:MATH:DUAL:SOURce<X> { CH1 | CH2 | CH3 | CH4 |
REF1 | REF2 | REF3 | REF4 | ? }
Parameter
<X>
Source number 1 or 2
CH1~4
Channel 1 to 4
REF1~4
Reference waveforms 1 to 4
Return parameter
Returns the source for the source 1 or 2.
Example
:MATH:DUAL:SOURce1 CH1
Sets source1 as channel 1.
:MATH:DUAL:OPERator
Description
Sets the math operator for dual math operations.
Set
Query
Set
Query
Set
Query