EasyManua.ls Logo

GW Instek GDS-1000B Series - Vertical Commands; Channel<X>:Bwlimit

GW Instek GDS-1000B Series
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDS-1000B Programming Manual
48
Vertical Commands
:CHANnel<X>:BWLimit ............................................... 48
:CHANnel<X>:COUPling............................................. 49
:CHANnel<X>:DESKew ............................................... 49
:CHANnel<X>:DISPlay ................................................ 49
:CHANnel<X>:EXPand ................................................ 50
:CHANnel<X>:IMPedance? ......................................... 50
:CHANnel<X>:INVert .................................................. 51
:CHANnel<X>:POSition .............................................. 51
:CHANnel<X>:PROBe:RATio ...................................... 52
:CHANnel<X>:PROBe:TYPe ........................................ 52
:CHANnel<X>:SCALe .................................................. 52
:CHANnel<X>:BWLimit
Description
Sets or returns the bandwidth limit on/off.
Syntax
:CHANnel<X>:BWLimit {FULL | <NR3> | ?}
Parameter
<X>
Channel 1,2,3,4
FULL
Full bandwidth
<NR3>
Sets the bandwidth limit to a pre-defined
bandwidth.
20E+6: 20MHz
Return Parameter
<NR3>
Returns the bandwidth.
Full
Full bandwidth
Example
:CHANnel1:BWLimit 2.000E+07
Sets the channel 1 bandwidth to 20MHz.
Set
Query

Table of Contents

Other manuals for GW Instek GDS-1000B Series

Related product manuals