COMMAND DETAILS
143
Related
commands
:TRIGger:BUS:B1:I2C:ADDRess:MODe
Parameter
<sting>
7/10 characters, must be enclosed in
double quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the address value.
Example1
:TRIGger:BUS:B1:I2C:ADDRess:MODe ADDR7
:TRIGger:BUS:B1:I2C:ADDRess:VALue "xxx0101"
Sets the address to XXX0101
Example 2
:TRIGger:BUS:B1:I2C:ADDRess:VALue?
XXX0101
:TRIGger:BUS:B1:I2C:ADDRess:DIRection
Description
Sets or queries the address bit as read, write or
don’t care.
Note
This setting only applies when the I
2
C trigger is set
to trigger on Address or Address/Data
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:I2C:ADDRess:DIRection { READ |
WRITE | NOCARE | ? }
Related
commands
:TRIGger:BUS:B1:I2C:CONDition
Parameter
READ
Set read as the data direction.
WRITE
Set write as the data direction.
NOCARE
Set either as the data direction.
Return Parameter
Returns the direction (READ, WRITE, NOCARE).
Set
Query