EasyManua.ls Logo

GW Instek GDS-1000B Series - Page 245

GW Instek GDS-1000B Series
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND DETAILS
245
Related
commands
:DATALOG:STATE
:DATALOG:SAVe
:DATALOG:INTerval
:DATALOG:DURation
Parameter/Return
parameter
CH1 ~CH4
Channel 1, 2, 3 or 4
all
All displayed channels.
Example
:DATALOG:SOURce CH1
Sets the source to CH1.
:DATALOG:SAVe
Description
Sets or queries the save format as image or
waveform.
Syntax
:DATALOG:SAVe {IMAGe|WAVEform|?}
Related
commands
:DATALOG:STATE
:DATALOG:SOURce
:DATALOG:INTerval
:DATALOG:DURation
Parameter/Return
parameter
IMAGe
Save as images.
WAVEform
Save as waveforms.
Example
:DATALOG:SAVe WAVEform
Sets the save format to waveform.
:DATALOG:INTerval
Description
Sets or queries the recording interval time in
seconds. The interval times that can be set are
dependent on the settings of the
DATALOG:SOURCE and DATALOG:SAVE
commands.
Syntax
:DATALOG:INTerval {<NR1>|?}
Set
Query
Set
Query

Table of Contents

Other manuals for GW Instek GDS-1000B Series

Related product manuals