REMOTE CONTROL
237
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:REAL[:HIGH]?
Returns the real current value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP1:AC:LIM:REAL?
> +2.000000E-03
The real current value for ACW of step1 set is 2mA.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:DWELl
Sets the wait time value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP1:AC:TIME:DWEL 0.5
Sets the wait time value for ACW of step1 to 0.5s.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:DWELl?
Returns the wait time value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP1:AC:TIME:DWEL?
> +5.000000E-01
The wait time value for ACW of step1 set is 0.5s.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:RAMP
Sets the ramp time value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>