[:SOURce]:SAFEty:RESult[:LAST]:MMETerage?
Returns the MEASURE METER readings of the last STEP.
Return Parameter: <NR3>measure meter
Example: SAFE:RES:LAST:MMET?
> +5.000000E-05
The MEASURE METER result is 0.05mA.
[:SOURce]:SAFEty:RESult:STEP<n>[:JUDGment]?
Returns the judgment results of selected STEP.
Parameter<n>: <NR1>(1 ~ 99)
Return Parameter: <NR1>result
Example: SAFE:RES:STEP2:JUDG?
> 116
The judgment result of the second STEP is PASS.
Common judgment result code list
Screen
Judgment Result
Code (Decimal)
PASS
PASS
116
STOP
STOP
113
Message
CAN NOT TEST
114
TEST
TESTING
115
STOP
STOP
112
GR CONT.
GR CONT.
120
GFCI
GFCI
121
POWERGND
POWER GND
122
V OVER
VOLT OVER
123
V LOW
VOLT LOW
124
Judgment result no good code list
Screen
Code Meaning
AC Mode
DC Mode
IR Mode
OSC Mode
HI SET
HI SET
17
33
49
---
LO SET
LO SET
18
34
50
---
ARC
ARC
19
35
---
---
SHORT
SHORT
---
---
---
97
OPEN
OPEN
---
---
---
98

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9500 Series and is the answer not in the manual?

GW Instek GPT-9500 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9500 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals